Electromagnetic Compatibility Information

1. This Alvita Infrared Ear Thermometer needs to be installed and put into service in accordance with the information
provided in the user manual.
2. WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of this device, including cables specified by
the manufacturer. Otherwise, degradation of the performance of this device could result.
If higher IMMUNITY TEST LEVELS than those specified in Table 9 are used, the minimum separation distance may
be lowered. Lower minimum separation distances shall be calculated using the equation specified in 8.10.
Manufacturer's declaration-electromagnetic immunity.
The Alvita Infrared Ear Thermometer is intended for use in the electromagnetic environment (for home healthcare)
specified below. The customer or the user of the Alvita Infrared Ear Thermometer should assure that is used in
such environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
More information on EMC compliance of the device may be obtained from Rossmax using the contacts shown in this
manual.
10
12367-MULTI INFRARED THERMOMETER BOOKLET ES, FR, DE, IT, ES, PT V1AW.indd 10
IEC 60601
Compliance
test level
level
3 Vrms:
Not applicable
0,15 MHz – 80 MHz
6 Vrms: in ISM and
Not applicable
amateur radio bands
between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
10 V/m
80 MHz – 2,7 GHz
80 MHz – 2,7 GHz
80 % AM at 1 kHz
80 % AM at 1 k
Electromagnetic environment-guidance (for
home healthcare environment).
Portable and mobile RF communications
equipment should
be used no closer to any part of the
Alvita Infrared Ear Thermometer including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance:
d = 1,2 √P, d = 1,2 √P 80MHz to
800 MHz, d = 2,3 √P 800MHz to 2,7 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Interference may occur in the vicinity of
equipment marked with the following
symbol: