1.3 Caution
Keep the oximeter away from dust, vibrations, corrosive substances, explosive materials, high temperatures
and moisture.
If the oximeter gets wet, stop using it immediately.
Do not use immediately after transfer from a cold environment to a warm or humid environment.
DO NOT operate the keys on the front panel using sharp materials.
High-temperature or high-pressure steam disinfection of the oximeter is not permitted. Refer to the rele-
vant chapter of the User Manual for cleaning and disinfection instructions.
Do not immerse the oximeter in a liquid. When cleaning, wipe its surface using medical-grade alcohol on a
soft cloth. Do not spray any liquid directly onto the device.
When cleaning the device with water, the temperature should not exceed 60°.
If the patient's fingers are too thin or too cold, this may affect the normal SpO
ments. In this case, clip a thicker finger such as the thumb or middle finger far enough into the probe.
Do not use the device on infants or neonatal patients.
The product is suitable for adults and children over four years old (Weight should be between 15kg and
110kg).
The device may not work on all patients. If you are unable to achieve stable readings, discontinue use.
The data is updated at intervals not exceeding 5 seconds; this may change according to an individual's
specific pulse rate.
The waveform is normalized. Read the measured value when the waveform shown on the screen is stable
and regular. This measured value is the optimal value and the given waveform is standard.
and pulse rate measure-
2
Oxystart® - EN
33