Guidance and manufacture's declaration-electromagnetic immunity
Guidance and manufacturer's declaration-electromagnetic immunity
The Oxystart
Pulse Oximeter is intended for use in the electromagnetic environment specified below. The
®
user of the Oxystart
Pulse Oximeter should ensure that it is used in such an environment.
®
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50Hz) magnetic
field
IEC 61000-4-8
50
Oxystart® - EN
for all EQUIPMENT and SYSTEMS
IEC60601 test
Compliance level
level
±6KV contact
± 6 kV contact
±8KV air
± 8 kV air
3 A/m
3 A/m
Electromagnetic
environment-guidance
Floors should be made of wood, concrete or
ceramic tile. If the floor is covered with syn-
thetic material, the relative humidity should
be at least 30%.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment