Setting Reflected Temperature Compensation
To enable and set reflected temperature compensation, use the main program menu
to select:
[Setup] → [Reflection Compensation]
Reflected temperature compensation can be enabled/disabled using [Compensate
for Reflected Temperature], and when enabled the temperature value can be set
between -40 and 1000°C.
operation
Once the sensor is in position and the appropriate power, air, water, and cable con-
nections are secure, the system is ready for continuous operation by completing the
following simple steps:
1. Turn on the power supply
2. Turn on the instrument or PC
3. Read / monitor the temperature
iMportant
Be aware of the following when using the sensor:
• If the sensor is exposed to significant changes in ambient temperature (hot to cold,
or cold to hot), allow 20 minutes for the temperature to stabilise before taking or
recording measurements.
• Do not operate the sensor near large electromagnetic fields (e.g. around arc welders
or induction heaters).
Electromagnetic interference can cause measurement errors.
• Wire must be connected only to the appropriate terminals.
maintenance
Our customer service representatives are available for application assistance, calibra-
tion, repair, and solutions to specific problems. Contact our Service Department before
returning any equipment. In many cases, problems can be solved over the telephone.
If the sensor is not performing as it should, try to match the symptom below to the
problem. If the table does not help, call Omega for further advice.
troubleshootinG
symptom
No output
Erroneous temperature
Erroneous temperature
Erroneous temperature
lens cleaninG
Keep the lens clean at all times. Any foreign matter on the lens would affect measure-
ment accuracy. Blow off loose particles (if not using the air purge accessory) with an
air 'puffer'.
probable cause
No power to sensor
Incorrect wire connection
Faulty sensor cable
Field of view obstruction
solution
Check power supply
Check wire colour codes
Verify cable continuity
Remove obstruction