Adjusting Display And Output Settings; Calibration Of Temperature Probes - CHART MVE TEC 3000 Guía De Referencia Rápida

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Adjusting Display And Output Settings

The units of measurement displayed by the TEC 3000 may be adjusted to accommodate the needs of the user.
Temperature measurement may be displayed in degrees Kelvin (°K), degrees Celsius (°C), or degrees Fahrenheit
(°F). The amount of liquid nitrogen in the freezer may be displayed in inches (in), millimeters (mm). In addition, the
amount of liquid nitrogen consumed by the freezer (liquid usage) may be shown on the display.
NOTE: Security Level 1 is required to adjust the display and output settings (See "Password and Security Setup").
section for details).
1. PRESS "SETUP"
The controller will prompt for a password. Use the
"
L
/ M" keys to scroll to the appropriate number. Press
L
/
"Enter" to move cursor to the next position in the password.
2. PRESS "SETUP"
Until the display reads "Display and Output".
3. PRESS "ENTER"
The display will read "Temperature Units". Use the
"
L
/ M" keys to toggle between °C (Celsius),
L
/
°F (Fahrenheit), or °K (Kelvin). Press "Enter" to save
settings.
4. PRESS "SETUP"
The display will read "Level Units". Use the "
to toggle between "in" (inches) or "mm" (millimeters).
Press "Enter" to save settings.
5. PRESS "SETUP"
The display will read " Display Liquid Usage". Use the
"
L
L
/ M" keys to select " Yes" or "No". Press " Enter" to
/
save settings.
After all adjustments are complete, wait 30 seconds and the TEC 3000 will return to the Main Display screen
indicating new settings are in effect

Calibration of Temperature Probes

Each of the temperature sensor probes (A and B) used with the TEC 3000 have been calibrated at the factory using
the "Low Temperature Range" method. This calibration method provides a level of accurancy of +/-1.8°F (+/-1°C)
when operated in the altitude range of 1000ft. to 1500ft. (305m to 457m). Further calibration should not be required
unless desired by the end user. Refer to the TEC 3000 Technical Manual for information on calibration methods
and procedures.
TEC 3000 Quick Reference Guide
.
Higher User Level
Required Use
Enter Password
0000
Press Enter For
Display And Output
Or Press Setup For
Next Menu
Temperature Units
°C
Use
Press Enter To Save
L
L
/ M
/
Level Units
In
Use
L
/ M" keys
L
/
Press Enter To Save
L
/ M
L
/
8
To
L
L
/ M
/
To Adjust
To Adjust

Publicidad

Capítulos

Tabla de contenido
loading

Productos relacionados para CHART MVE TEC 3000

Este manual también es adecuado para:

Mve serie

Tabla de contenido