12. Troubleshooting

Fault
Cause
Device shows no
Incorrect mains
function
connection
Main switch off
Internal cabling is
defective
Plug-in power supply
unit is defective
LED lights up, electronics
Electronics are defective
without function
Rotary plate without
Function deactivated
function
Device cover is open
Cable connection to
rotary motor (16) is
defective
Rotary motor (16) is
defective
Electronic module is
defective
Motor adapter (25)
rotates continuously
Time programmes can
Main switch off
not be activated
Device cover is open
Internal cabling is
defective
26
MAX
Troubleshooting
Check mains connection
Switch on main switch
Internal cabling must be
checked by a specialist.
Replace the plug-in
power supply unit
Electronics must be
checked by a specialist
Activate with pressure
switch (5)
Close cover (12)
Examine the cabling
The motor must be
examined by a special-
ist, and replaced if
necessary
The electronics must be
examined by a special-
ist, and replaced if
necessary
Tighten the motor
adapter threaded pin
Switch on the device
Close cover (12)
Internal cabling must be
checked by a specialist
Fault
Cause
LED circuit board (9) is
defective
Electronic module is
defective
Micro-switch (15) is
defective
Polymerisation time
Electronic module is
does not end automati-
defective
cally
Polymerisation process
Both micro-switches
(15) are defective
is not interrupted when
the cover is opened
Unsatisfactory polymeri-
Polymerisation time is
sation results
too short
Material reacts to
fluctuating wavelengths
LED circuit board (9) is
soiled
LED circuit board is
damaged
Troubleshooting
The LED circuit board
must be examined by a
specialist, and replaced
if necessary
The electronics must
be examined by a
specialist, and replaced
if necessary
The micro-switch must
be checked by a spe-
cialist and replaced if
necessary
The electronics must
be examined by a
specialist, and replaced
if necessary
Deactivate the
device immediately.
The micro-switch must
be checked by a
specialist and replaced
if necessary
Extend the polymerisa-
tion time
Replace the material
Clean the LED circuit
board carefully
Examine the LED
circuit board, replacing
if necessary
27