8
Maintenance
The unit must be unplugged (6) before it is cleaned. Use a moist cloth to clean the unit and make sure that no water
penetrates into the unit. Do not use aggressive cleaning agents (scouring powder, solvents).
The unit consists of Makrolon components which may only be cleaned with mild cleaning agents or ethanol (96%).
Spare fuses can be purchased from Kulzer or specialized dealers and may be exchanged by the customer.
Attention!
The unit must be unplugged before exchanging fuses!
9
Technical data
GB
Power consumption
- 10 -
Rated voltage / fuse type
Nominal frequency
Protective class
Pollution degree
Pressure hose connection
Nominal pressure
Max. operating pressure
Operating temperature of water
Height
Width
Depth
Weight
10
Information on malfunctions
Error messages
Error
Cause
E 98
Breakage / short circuit of sensor (heating
from software
plate)
version C 14
or
Automatic rapid deactivation (software) owing
to excess temperature.
E 99
Breakage / short circuit of sensor (water bath)
or
Automatic rapid deactivation (software) owing
to excess temperature.
E 01
Process start attempted without timer.
E 02
Process start attempted without (previous)
heating function.
650 W
100 V, 2 x T8A / 250 V
120 V, 2 x T6.3A / 250 V
230 / 240 V, 2 x T4A / 250 V
50 / 60 Hz
1
2
3 to 10 bars
2 bars
2.8 bars
45°C (113°F) / 55°C (131°F) / 100°C (212°F)
deviation +/– 2.5°C (4.5°F) or approx. 5%
Dimensions
Required space
approx. 340 mm
approx. 500 mm
approx. 360 mm
approx. 370 mm
approx. 350 mm
approx. 500 mm
approx. 6 kg
Remedy
•
Switch off the unit.
•
Check water filling level (see chapter 7.2 Filling with water).
•
Allow the unit to cool down approx. 10 min.
•
If the error can not be eliminated, inform your responsible service agent
(see chapter 11 Service).
•
Switch off the unit.
•
Check water filling level (see chapter 7.2 Filling with water).
•
Allow the unit to cool down approx. 10 min.
•
If the error can not be eliminated, inform your responsible service agent
(see chapter 11 Service).
•
Set process timer (see chapter 7 Working with the Palamat elite).
•
Activate standby heating function (see chapter 7.3.3 Starting a process).
•
If the error can not be eliminated, inform your responsible service agent
(see chapter 11 Service)