ENGLISH INSTRUCTIONS

Only a certified company or
electrician should install the cabinet
since part of the installation comes
with an electrical installation. Your
cabinet is not yet ready for use!
To ensure the defibrillator stays
within its operational temperature
range during the whole year,
the Rotaid should be located in
the shade at all times.
Step 1. Preparations and drilling
The ideal mounting position for the Rotaid
150
is 150 centimeters (59 inches), measured
CM
from the center of the cabinet to the floor.
Use the drilling-template to mark the four holes for the
mounting positions. Also mark the hole for the cable to
the cabinet.
Please drill the four holes using an 8mm drill
and a 10mm drill for drilling the cable hole.
Now place the four plastic wall plugs.
Step 2. Cable feed and fixation
Please find assistance from another person to assist you
with this step.
On the bottom-left of the backplate a rubber plug is
located. Use an awl or screwdriver to puncture this plug to
allow for enough space for the power cable.
4.
WARNING
Run the whole power cable through the hole and wall.
Important! Do not yet attach the white Rotaid 24/7
module to the back plate.
Now attach the back plate to the wall by using the four
screws.
After this you can mount the 24/7 module via the two
small screws. Please use the two holes on the top of the
module (see picture).
Determine the location where you
wish to mount the Rotaid.
Make sure there are no irregularities
present at the desired location.
Please take into account the fire safety
and possible evacuation regulations.