Guidance And Manufacturer's Declaration -Electromagnetic Immunity - Pari Vios PRO Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

K. SPECIFICATIONS: IEC TABLES
The Following Tables Are Provided In Accordance With IEC 60601-1-2:
between portable and mobile RF communications equipment and the
The Vios
is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
TM
or the user of the Vios
mobile RF communications equipment (transmitters) and the Vios
a minimum as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
The Vios
is intended for use in the electromagnetic environment specified below. The customer or the user of the Vios
TM
should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
EN-9
Questions? 1.800.FAST.NEB (327.8632) | PARI.com/VIOSPRO
Recommended separation distances
can help prevent electromagnetic interference by maintaining a minimum distance between portable and
TM
150 kHz to 80 MHz
W
d =
0,01
0,1
1
10
100
Guidance and manufacturer's declaration –electromagnetic immunity
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 sec
3 A / m
can help prevent electromagnetic interference by maintaining
TM
Separation distance according to frequency of transmitter
80 MHz to 800 MHz
3,5
P
d =
V
1
.12
.12
.37
.37
1.17
1.17
3.69
3.69
11.67
11.67
Compliance level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 sec
3 A / m
Vios
TM
m
800 MHz to 2,5 GHz
3,5
P
d =
E
E
1
.23
.74
2.33
7.38
23.33
Electromagnetic environment
guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Vios
requires continued
TM
operation during power mains
interruptions, it is recommended
that the Vios
be powered from
TM
an uninterruptible power supply
or a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
7
P
1
TM

Publicidad

Tabla de contenido
loading

Tabla de contenido