6.3 Viscosimeter calibration
This menu is used for viscosity and temperature calibration.
IMPORTANT:
a) The viscosimeter comes factory calibrated with the corresponding certificate.
b) Periodical calibration of the viscosimeter is essential to get reliable and traceable
measurements, but if this calibration is not properly done, it could cause a large error
in measurements.
c) The equipment is ready to be calibrated by the user, but you must have the
necessary material and have the adequate knowledge on viscosity measurement.
d) J.P. SELECTA, S.A. is not liable for any measurements taken once the equipment
has been recalibrated by the user.
e) «Factory» calibration can always be recovered. This is useful when we get
questionable results after a calibration.
Materials needed for viscosimeter calibration:
a ) Standard oils:
They should be silicone oils of the following viscosities:
Table 6.1:
Model:
L
R
b ) Standard
calibration.
For th e te mperatu re mea su re me nt ca librat ion, you in it ia lly nee d a ca librat ed
thermometer to compare the temperature measured by the viscosimeter with the one
indicated in the thermometer.
Considering that the difference between both is unacceptable, you must adjust the
measurement circle. To do this, you need a pT100 probe simulator at temperatures of
0ºC, 30ºC and 100ºC.
Factory calibration recovery (Reset):
Make a viscosimeter «reset». This means that the viscosimeter recovers the original
factory calibration, in terms of viscosity and temperature.
Select RESET (Fig 6.6) and the RESET message will appear (Fig 6.7).
To proceed with the process, press «ENTER». Otherwise, press «QUIT».
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
Febrero/2011
(It can be modified without notice)
Spindle:
Standard oil type:
L1
RT50
L2
RT500
L3
RT1000
L4
RT5000
R1
RT50
R2
RT500
R3
RT500
R4
RT1000
R5
RT5000
R6
RT5000
R7
RT30000
thermometer
and
«gauge»
fo r
temperature