3B SCIENTIFIC PHYSICS 1000620 Instrucciones De Manejo página 13

Tubo del potenciales críticos s, llenado de helio
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 33
+
1
2
RING
Fig. 4 Connection of an oscilloscope to the control unit for critical potential tubes
7.3 Experiment set-up with the control unit
for the Franck-Hertz experiment
Insert the tube into the tube holder.
Provision of heater voltage V
Connect socket F3 on the tube holder to
socket F on the control unit for the Franck-
Hertz experiment and connect socket F4 to
socket K. (refer to Fig. 5)
Provision of accelerating voltage V
Connect socket C5 on the tube holder to
socket K on the control unit and socket A1 to
socket A.
Fig. 5 Experiment set-up with the control unit for the Franck-Hertz experiment
7.3.1
Procedure with 3B NETlog
Connect the 3B NETlog
unit for the Franck-Hertz experiment (see
Fig. 6). Connect the Ux output of the control
unit to input A of the 3B NETlog
output Uy to input B.
Set the control unit to ramp mode and con-
figure a minimum voltage of about 10 V, with
a maximum voltage of about 35 V.
Select a heater voltage of some 3.5 V and a
collector voltage of around -1.5 V.
Set up the 3B NETlog
TM
3B NETlab
computer program and the
OUT
3
60
200 mV
!
V
A
4
HERTZ TUBE CONSOLE
0...±1 VOLT OUT
0...±1 VOLT OUT
SLOW
FAST
1
2
1
:
F
:
A
TM
TM
unit to the control
TM
unit and
TM
interface, the
OSCILLOSCOPE
MAX
3
V
SET
A
4
MIN
SLOW
2
RUN
CH1
(X)
The provision of collector voltage V
internally inside the control unit for the Franck-
Hertz experiment.
Put the shielding over the tube and slot it into
the groove on the tube holder so that the tube
is completely enclosed by shielding. Then
connect the control unit by means of the
ground socket.
Connect the collector ring to the BNC input
on the control unit.
BETRIEBSGERÄT FRANCK-HERTZ/OPERATING UNIT FRANCK-HERTZ
Heizung
Filament
C5
F4
U
F
V
F3
F
A1
K
measurement recording as described in sec-
tion 7.2.1.
Vary the heating voltage, the upper and lower
limits of the accelerating voltage and the gain
until the best optimal can be seen.
In order to display the ionisation curve, re-
verse the polarity of the collector voltage.
7.3.2
Procedure with an oscilloscope
Connect the Ux output of the control unit to
Channel 1 (X deflection) of the oscilloscope
and output Uy to Channel 2 (Y deflection)
(see Fig. 7).
5
CH2
EXT
(Y)
R
Gitter/Grid
Beschleunigung/Acceleration
Reverse bias
0
12
0
80
0
80
0
U
U
U
U
G
Amin
Amax
V
V
Man/Ramp
V
V
E
1
10
1
G
A
=
+
U
U
x
A
10
is handled
12
E
+/-
V
= *
U
I V
Y
E

Publicidad

Tabla de contenido
loading

Tabla de contenido