Data Transmission - Oregon Scientific RGR126 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 15
2.
Set the channel switch to any channel. The switch
is located in the battery compartment.
3.
Press RESET on the sensor.
4.
Close the remote sensor battery compartment.
5.
Secure the sensor in the desired location using the
wall mount or table stand.
For best results:
Insert the batteries before you mount the sensor.
Place the sensor out of direct sunlight and
moisture.
Do not place the sensor more than 00 meters
(328 feet) from the main (indoor) unit.
Position the sensor so that it faces the main
(indoor) unit, minimizing obstructions such as
doors, walls, and furniture.
Place the sensor in a location with a clear view to
the sky, away from metallic or electronic objects.
Position the sensor close to the main unit during
cold winter months as below-freezing temperatures
may affect battery performance and signal
transmission.
You may need to experiment with various locations to
get the best results.
Standard Alkaline batteries contain significant amounts
of water. Because of this they will freeze in low
temperatures of approximately -2°C (0°F). Disposable
Lithium batteries have a much lower threshold for
temperature with an estimated freezing range of below
-40°C (-40°F).
Wireless ranges can be impacted by a variety of factors
such as extremely cold temperatures. Extreme cold
may temporarily reduce the effective range between the
sensor and the base station. If the unit's performance
fails due to low temperature, the unit will resume proper
functioning as the temperature rises to within the normal
temperature range (i.e. no permanent damage will occur
to the unit due to low temperatures).

DATA TRANSMISSION

Data is sent from the remote thermo sensor approx.
every 78 seconds and every 94 seconds from the
rain gauge. The reception icon shows the status - see
table.
7

Publicidad

Tabla de contenido
loading

Tabla de contenido