Gemini-20
6. Once the weight is accepted, the display will show
"PASS" then return to normal weighing. Calibration
is complete!
IX. Troubleshooting
Problem: Scale will not turn on
Solution: Check the batteries
Problem: Displayed weight fluctuates randomly
Solution: Make sure your workspace is stable and
free of any air currents or vibration
Problem: Displayed weight is inaccurate
Solution: Perform a calibration adjustment
X. Error Codes
1. OUTZ - Zero range has shifted. Perform calibration
adjustment. If the problem persists, the load cell may
be permanently damaged due to overloading.
2. 88888 - Maximum capacity exceeded. Remove the
extra weight to avoid damaging the load cell.
3. L - Batteries are low. Replace the batteries.
Max. Capacity
Readability
Power
Dimensions
Platform Dimensions
Warranty
- English
10 Year Limited Warranty
Gemini-20
20g
0.001g
2 x AAA
3.0 x 5.0 x 0.75"
1.5" Ø