Troubleshooting

Problem
The display does not come on & there
is no signal from the unit
The unit is switched on but does not
respond to commands
Battery symbol flashing: Ineffective
stimulation
Stimulation received irregularly, only
at a high intensity, or not at all
Increasing intensity causes
unpleasant sensation
Load Sense symbol appears
Load Sense symbol appears
beeping
14
Possible Cause
Battery discharged
Battery was incorrectly positioned
Lead not fully inserted
Broken lead
The battery is low
Faulty lead
Check your skin for lotions, pigment marks,
dry marks or other factors that could increase
resistance.
, unit
, unit
Faulty lead assembly
Electrode faulty
Poor skin/electrode contact
Solution
Replace battery
Remove battery, replace correctly
Remove plug, reinsert
Replace electrode/ lead assembly
Replace the battery
Replace lead
Slowly move electrode to locate area
where stimulus feels strongest
Moisten electrodes
Wash any oils from the skin
Check connections, replace if broken
Replace electrode
Check electrode contact with skin