I-Tech Mio-Sonic Manual De Uso página 20

Ultrasonidos
Ocultar thumbs Ver también para Mio-Sonic:
Tabla de contenido

Publicidad

EMC Compliance
Emission Test
RF Emissions
Cispr 11
RF Emissions
Cispr 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations /
Flicker emissions
IEC 61000-3-3
Mio-Sonic is intended for use in the electromagnetic environment specified
below. The customer or the user should assure that it is used in such an
environment.
Immunity Test
Electrostatic
Discharge (ESD)
EN 61000-4-2
Electrical fast
transient /burst
EN 61000-4-4
Surge
EN 61000-4-5
IACER Srl
Electromagnetic emission
Compliance
Group 1
Class B
Class A
Complies
Complies
Electromagnetic immunity
EN 60601-1-2
Test level
± 6kV contact
± 6kV contact
± 8kV air
± 8kV air
±2kV for
±2kV for power
power supply
supply lines
lines
±1kV
±1kV
differential
differential
mode
mode
20 di 24
Electromagnetic environment
- guidance
Mio-Sonic device uses RF
Energy only for its internal
function. Therefore its RF
emission are very low and are
not likely to cause any
interference in nearby
electronic equipment.
Mio-Sonic device is suitable
for use in all establishments,
including domestic
establishments and those
directly connected to the public
low-voltage power supply
network that supplies buildings
used for domestic purposes.
Compliance
Level
Floors should be wood,
concrete or ceramic tile.
If floor are covered with
synthetic material, the
relative humidity should
be at least 30%.
Mains
should be that of a
typical commercial or
hospital environment.
Mains
should be that of a
typical commercial or
hospital environment.
MNPG72 Rev. 02 del 01/03/15
Electromagnetic
environment -
guidance
power
quality
power
quality

Publicidad