5.0 Cable Management
Two (2) Cable Guides are provided to facilitate routing of cables along the underside of the Arm. Squeeze the Cable
Guide until its edges snap into the grooves in the bottom surface of the Arm. Leave some extra cable loose at the front
and rear of the arm to prevent cable binding, connector damage, or Cable Guide damage.
6.0 Routine Maintenance
Periodically inspect all tilting, swiveling, pivoting, and mounting hardware. Tighten or adjust as necessary for optimal
operation and safety.
7.0 Cleaning the Mounting Assembly
7.1. The mounting assembly may be cleaned with most mild, non-abrasive solutions commonly used in the hospital
environment (e.g. diluted bleach, ammonia, or alcohol solutions).
7.2. The surface finish will be permanently damaged by strong chemicals and solvents such as acetone and
trichloroethylene.
7.3. Steel wool or other abrasive material should never be used.
7.4. Damage caused by the use of unapproved substances or processes will not be warranted. We recommend testing
any cleaning solution on a small area of the arm that is not visible, to verify compatibility.
7.5. Never submerge or allow liquids to enter the arm. Wipe any cleaning agents off of the arm immediately using a
water-dampened cloth. Dry the arm thoroughly after cleaning.
CAUTION: GCX makes no claims regarding the efficacy of the listed chemicals or processes as a means for controlling
infection. Consult your hospital's infection control officer or epidemiologist. To clean or sterilize mounted instruments or
accessory equipment, refer to the specific instructions delivered with those products.
Cable Guides (2)
DU-MND–0013-01 Rev A 9/28/10 GCX Corp. Page 9 of 10