60.00 kΩ
600.0 kΩ
6.000 MΩ
60.00 MΩ
Accuracy (cont'd)
99.99 nS
Continuity
Tone Threshold: Between 20 Ω and 300 Ω
Continuity
Response Time: < 100 μs
Tone Threshold: Between 20 Ω and 300 Ω
Response Time: < 100 µs
Capacitance
Range
60.00 nF
600.0 nF
6.000 µF
60.00 µF
600.0 µF
(2)
6.000 mF
(2)
(2)
25.00 mF
(1)
Accuracies with film capacitor or better
(2)
In manual ranging mode, measurements are not specified below 50.0 µF, 0.54 mF, and 5.4 mF for
600.0 µF, 6.000 mF and 25.00 mF ranges, respectively
Diode Test
Measuring Range: 2.000 V
Test Current (typical): 0.4 mA
Accuracy (cont'd)
Open Circuit Voltage: < 3.5 VDC
AC Current
Accuracy: 1.0% + 0.001 V
AC Current (AC + DC Current on DM-830A and DML-430A Only)
Accuracy at
Range
50 to 60 Hz
600.0 µA
± (0.6% + 0.3 µA)
6000 µA
± (0.6% + 3 µA)
60.00 mA
± (0.6% + 0.03 mA)
600.0 mA
± (1.0% + 0.3 mA)
6.000 A
± (0.8% + 0.006 A)
10.00 A
± (0.8% + 0.06 A)
10 A continuous, 20 A for 30 sec. maximum with 5 minute cool down.
DC Current
Range
600.0 µA
6000 µA
60.00 mA
600.0 mA
6.000 A
10.00 A
10 A continuous, 15 A for 30 sec. maximum (20 A for DML-430A) with 5 minute cool down.
10 A continuous, 15 A for 30 sec. maximum with 5 minute cool down.
Line Level Frequency
Function
60.00 mV
600.0 mV
± (0.1% + 0.03 kΩ)
± (0.1% + 0.3 kΩ)
± (0.4% + 0.003 MΩ)
± (1.5% + 0.05 MΩ)
± (0.8% + 0.1 nS)
Accuracy
(1)
± (0.8% + 0.03 nF)
± (0.8% + 0.3 nF)
± (1.0% + 0.003 µF)
± (2.0% + 0.03 µF)
± (3.5% + 0.5 µF)
± (5.0% + 0.005 mF)
± (6.5% + 0.05 mF)
Accuracy at
40 to 1 kHz
± (0.8% + 0.4 µA)
± (0.8% + 4 µA)
± (0.8% + 0.04 mA)
± (1.0% + 0.4 mA)
± (0.8% + 0.006 A)
± (0.8% + 0.06 A)
Accuracy at 50 to 60 Hz
± (0.2% + 0.4 µA)
± (0.2% + 4 µA)
± (0.2% + 0.04 mA)
± (0.2% + 0.4 mA)
± (0.2% + 0.004 A
± (0.2% + 0.04 A)
Sensitivity (Sine RMS)
15
40 mV
60 mV
1.2 VDC
1.2 VDC
Burden Voltage (typical)
(all frequency ranges)
0.08 mV/µA
2.1 mV/mA
0.02 V/A
Burden Voltage (typical)
(all frequency ranges)
0.08 mV/µA
2.1 mV/mA
0.02 V/A
Range
15.00 Hz to 50.00 kHz
15.00 Hz to 50.00 kHz
MM810A
21