ERROR CONDITIONS
DESVIATION OF TEMPERATURE
The default working temperature is 4ºC ±2ºC in Blood Bank C/D and
20ºC ±2ºC in Blood Bank E/F. If the temperature rises outside the wor-
king range, either above or below, an error occurs which is displayed
on the screen (See previous table).
Both the working temperature and the upper and lower limits are pro-
grammable as explained in the section "Modification of parameters".
NOTES ON ALARMS
- The acoustic signals that cause the error conditions can be stopped
by pressing the ON/STAND-BY key.
- There is the possibility of using one of the board relays to activate
an external alarm in the event that the error condition occurs (See
connection diagrams).