Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

MANUAL DE INSTRUCCIONES CODIGO 80382
J
.P. SELECTA s.a.u.
REV A

ERROR CONDITIONS

DESVIATION OF TEMPERATURE
The default working temperature is 4ºC ±2ºC in Blood Bank C/D and
20ºC ±2ºC in Blood Bank E/F. If the temperature rises outside the wor-
king range, either above or below, an error occurs which is displayed
on the screen (See previous table).
Both the working temperature and the upper and lower limits are pro-
grammable as explained in the section "Modification of parameters".
NOTES ON ALARMS
- The acoustic signals that cause the error conditions can be stopped
by pressing the ON/STAND-BY key.
- There is the possibility of using one of the board relays to activate
an external alarm in the event that the error condition occurs (See
connection diagrams).
GUARANTEE
This product is guaranteed for one year. The guarantee does not cover
damages caused by an incorrect use or causes beyond the control
of J.P. SELECTA, s.a.u.
Any manipulation of the equipment by unauthorized personnel by J.P.
SELECTA, s.a.u. cancels the guarantee automatically.
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: www.jpselecta.es
04/2020
Tel 34 937 700 877 Fax 34 937 702 362
Pag.: 12

Publicidad

Tabla de contenido
loading

Tabla de contenido