Equipment Description - J.P. SELECTA BLOOD-BANK C Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

INSTRUCTIONS MANUAL
Part
Litres
Bags
capac.
capac.
number
2101512
600
200
2101514
1300
480
2101511
600
200
2101513
1300
480
J
.P. SELECTA s.a.u.
CODE 80382 REV A
Internal dimensions cm
Height
Length
Width
136
58
69
136
128
69
136
58
69
136
128
69

EQUIPMENT DESCRIPTION

BLOOD-BANK C/D storage cabinets are designed to work at fixed
temperatures at +4ºC (Stability ±1,5ºC) and BLOOD-BANK E/F
platelet incubators are designed for + 20°C fixed temperatures.
Equipped with an hermetic compressor mounted on shock absorbers
with evaporator and forced air condenser, air cooling system which
stops on opening the door, temperature sensors housed in similar
conditions to the blood bag, inner LED light, double insulated anti-mist
glass door and a circular chart recorder.
Stainless steel AISI 304 external cabinet. Inner lining also in stainless
steel. Double glazed antimist, reversible lockable door, easy to change,
self close door it left ajar.
Fan air circulation in order to reach homogeneous temperature
distribution. External alarm ability. Circular graphic register. Pt100
probe. Safety thermostat. Temperature controlled by microprocessor
which thought display allows view and control following parameters:
- Security password
- Set temperature
- Set alarms
- Calibration
- Start up cooling motor delay.
INSTALLATION
Avoid installing the equipment near heat sources (sunny places,
heating, etc). Leave a space of about 10 cm between the wall and
the back of the cabinet. Place the equipment on a hard horizontal
leveled floor; if it is necessary use the adjustable foot. Do not start
up before 4 hours.
VERY IMPORTANT BEFORE THE FIRST STARTING UP
Connect the equipment at mains.
The default temperature set is 4ºC (20ºC in the case of models E/F).
Do not fill the equipment if it has not reached this temperature.
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: www.jpselecta.es
External dimensions cm
Height
Length
Width
207
70
82
207
140
82
207
70
82
207
140
82
04/2020
Consum
Power
(W)
(HP)
shelves
350
3/8
5
1100
3/8
10
350
3/8
5
1100
3/8
10
Tel 34 937 700 877 Fax 34 937 702 362
Pag.: 9
Weight