Loading the probe
1.
4.
1. Accurate measurement is guaranteed only when using probes made by Icare Finland Oy.
2. Open the package, and load the probe into the probe base, according to the illustrations.
3. Place the opened disposable package against the collar component.
4. Turn the device and the probe package at the same time, so that the probe package is uppermost. Check that
the probe has dropped into the probe base before removing the package.
5. The probe in the probe base.
6. Raise the device to the vertical operating position ensuring the probe cannot fall out of the probe base.
Press the measurement button. The probe will be magnetized by moving back and forward for a short time,
and the mechanism that prevents the probe dropping out accidentaly will start.
The device is ready for measurement when the reading 00 appears in the display. If the display shows
the message LoAd, there is either no probe in the device, or the probe is stuck. Check that the probe
has not dropped out, or press the function button again.
Turning the tonometer OFF
1.
1. Press either of the selector buttons until the display shows End.
2. Press the measurement button for two seconds. The display shows ByE and the tonometer turns off. The used
probe is ejected from the tonometer. Be careful to dispose of the probe properly (p. 4).
3. The tonometer turns off automatically after two minutes idle time.
6
2.
5.
2.
3.
6.
3.