MANUAL DE INSTRUCCIONES
LANGUAGE
DATE/TIME
INSERT
CONTRAST
PASSWORD
INFO
LOGS DOWNLOAD
HOME
Fig 7.2 Setting screen
J
.P. SELECTA s.a.u. - Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
CODIGO 80296 REV A
7.2 Setting screen
UNITS
7.2.1 Set time and date
CHANGE
Press on Date/time button. Press successively on the box to select the day,
PASSWORD
month, year, hour, minute and second settings.
Use Up and Down icons to entry a value.
CAL
7.2.2 Set temperature, language and screen contrast
In a similar way, set the temperature, language and screen contrast by pres-
sing the icons.
7.2.3 Working parameters protected under password
There are several parameters under password "0000" which affect the equi-
pment operation.
. Press "Insert password" and follow the instructions:
. Press CAL to access these parameters:
a) Log interval
It is the frequency of data collection for the logging.
b) External sensor
Select the value YES if you want to work with an external probe connected
to the connector on the back side of the equipment
c) Calibration (OFFSET)
Use only this setting if you notice a deviation between the temperature
indicated by the equipment and the one indicated by a standard thermometer.
For example: the equipment indicates 40.0ºC, the pattern indicates 40.5ºC.
Enter a value of -0.5ºC to correct the deviation.
d) The lower and the upper limit of temperature
These are the maximum and the minimum limits of temperature ad-
justments. This allows you to adjust the working temperature range of the
equipment.
e) To update the firmware of the device, click on "Info" and the UPGRD
button will appear at the bottom of the screen.
Contact the technical service to obtain the updated file.
7.3
Up to 10 programms of temperature and time can be stored.
Each program has the following parameters:
Temperature:
Time:
Undertemp:
Overtemp:
From main menu, press over «PROGRAMMING».
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
Mar2017
(Sujetas a modificaciones sin previo aviso)
password:
log interval:
external sensor:
calibration:
lower temp. limit:
upper temp. limit:
Programming screen
from ambient +5ºC ... 200ºC
HH:MM
Alarm value for temperature drop.
Alarm value for overtemperature.
0000
1 to 60min
yes/no
-100 to +100
-10
-10
Pag.: 17