7. Error messages
P.FAIL Power failure. The controller continues with the
process.
Err
will appear in green on the top line of the display.
TSEG
Safety thermostat is open.
SENS
Open temperature probe. Control does not detect the
temperature probe.
HEAT
The heating element does not work. The equipment detects
that the temperature does not increase.
8. Settings
The offset (A) and the ºC/ºF unit (B) can be adjusted.
A. Adjust the probe reading correction (offset)
Press keys (2) and (3) continuously and at the same time (about 5
seconds). The OFFs message will appear.
Use keys (2) / (3) to set the actual temperature reading value. Press
key (4) to confirm.
B. Presentation unit: ºC / ºF
Press key (1). The message Unit will appear. Using keys (2) / (3), select
the unit: ºC / ºF. Confirm with key (4).
9. Maintenance
Before removing the casing disconnect the equipment from the mains.
The manipulation of the internal electronic circuits of the
equipment by unauthorized personnel can cause irreparable
damage. Take it to one of the J.P. SELECTA, s.a.u. authorized
technical services.
CLEANING:
For the cleaning of the different parts, use the following products:
Cleaning of stainless steel: Alcohol.
Cleaning of plastic: Alcohol with cotton duster.
10. Guarantee
This product is guaranteed for one year. The guarantee does not cover
damages caused by incorrect use or causes beyond the control of J.P.
SELECTA, s.a.u.
Any manipulation of the equipment by personnel not authorized by J.P.
SELECTA, s.a.u. automatically cancels the guarantee.