4126-120-700 Rev-L
Circulating Assistant Action
CAUTION: DO NOT contaminate the aseptic
housing. The transfer shield guards the
aseptic housing from contamination during
battery installation.
4. Hold the non-sterile battery by the handle.
5. While aligning the battery's housing profile
with the transfer shield and chamber
opening, place the non-sterile battery into
the chamber (Figure 3).
Figure 3 – Install Battery
6. Remove the contaminated transfer shield
and set it aside.
www.stryker.com
Print Date: Jun 14, 2016 11:18:38 AM
0000003946, Rev. L Effective Date: Jun 14, 2016 10:36:00 AM
WARNING: Sterilize the contaminated
transfer shield before using it again.
See the care instructions manual
supplied with the handpiece.
Sterile Assistant Action
7. Close the aseptic housing lid and rotate its
latch clockwise 90° until the detent clicks
into place. Ensure the marker aligns with
the lock symbol (Figure 4).
Figure 4 – Lock Lid
WARNING: Failure to turn the latch
completely to the lock symbol may
result in the lid opening, exposing the
non-sterile battery.
8. Slide the aseptic housing containing the
installed non-sterile battery firmly into the
handpiece until it snaps into place.
EN
7