Display: (Refer to Figure A.)
A)
This Green LED is illuminated whenever power is being applied to the heater in the Lava Bed
Humidifier Cartridge.
B)
This Green LED is illuminated when the monitored temperature (Lava Bed Humidifier Cartridge or
Airway) is within 1.5 °C of the temperature set on the Temperature Control. This Yellow LED is
illuminated when the monitored temperature (Lava Bed Humidifier Cartridge or Airway) is 1.5°C or
more below the temperature set on the Temperature Control.
C)
This Green LED is illuminated when the monitored temperature (Lava Bed Humidifier Cartridge or
Airway) is within 1.5 °C of the temperature set on the Temperature Control.
(D)
This Yellow LED is illuminated when the monitored temperature (Lava Bed Humidifier Cartridge or
Airway) is 1.5°C or more above the temperature set on the Temperature Control.

Cleaning and Maintenance:

Periodically the NeoPod T controller may be wiped off using a damp cloth.
Warning!
Unplug the cable from the NeoPod T prior to cleaning. Do not put the Controller in water or other
liquid as this may cause damage.

Servicing:

The NeoPod T should be inspected by an authorized Westmed Inc. repair specialist five years from the date of
manufacturing. Prior to this, the device is intended to provide safe and reliable operation provided that it is operated
and maintained in accordance with the instructions provided by Westmed Inc. As with all electrical devices, if any
irregularity becomes apparent, you should take caution and have the device inspected by a Westmed Inc. repair
specialist.

Specifications:

 Input Voltage: 12VDC +2.5 / - 1.5
Input Current: 1.7A (maximum at 10 Liters / Minute
and 38°C.)
Output Temperature Range: 30 – 38°C @100% RH
Maximum output temperature: 40°C.
Minimum Humidification output: 33mg H 2 O /
Liters/Minute
Maximum Humidification output: 44mg H 2 O /
Liters/Minute
 Input Flow rate: 0 to 10 Liters per minute @ 50psi
max
 Compliance: 0.66 ml/kPa full – 1.1ml/kPa empty
 Flow Resistance: 1 cm of H 2 O at 20 Liters / Minute
 Total System Leakage: less than 60ml/min at 60 cm
of H 2 O
Water reservoir capacity: 20ml
 Warm up time: 15 min maximum
Maximum system operating pressure: 70cm H 2 O
 Operating Temperature Range: 21-38°C
Gas Inlet Temperature Range: 21-38°C