Warranty:
All Westmed products are warranted to be free of defects in material and workmanship for a period of ninety (90)
days from date of purchase.
Classification according to IEC 60601-1
Classification of protection: class II
Powered by external 12 volt DC power supply
No internal power supply
No protective earth ground
Type of applied part: BF
Protection from ingress of water: none IPXØ
Mode of operation: capable of continuous operation
No known potential for adverse affects from electromagnetic or other interference between this device and
other known devices
No known adverse effects on the performance of the humidification system when exposed to, electrocautery,
electrosurgery, defibrillation, X-ray (gamma radiation), infrared radiation, conducted transient magnetic
fields including magnetic resonance imaging (MRI), and radiofrequency interference.
Gross Dimensions of Controller:
Weight – 180 grams
Height – 83 mm
Length – 118 mm
Width – 43 mm
Guidance and manufacturer's declaration - electromagnetic emissions
The EQUIPMENT is intended for use in the electromagnetic environment specified below.
The customer or the user of the EQUIPMENT should assure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage Fluctuations/
Flicker emissions
The EQUIPMENT is suitable for use in all establishments other than domestic and those directly connected to the
public low-voltage power supply network that supplies buildings used for domestic purposes.
Compliance
Group 1
Class A
Not Applicable
Not Applicable
The EQUIPMENT uses RF energy only for it internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
6