A unique binaural group ID programmed to the
processor and its accessories at fitting time for SBP
which is validated during each transmission.
•
A unique pairing key established at pairing time for
DM devices which is validated during transmission
•
A cyclic redundancy check (CRC) which ensures
the integrity of remote-control and audio data.
•
A system architecture that isolates the processor's
critical functionality of stimulation parameters from
the wireless functionality
Bluetooth safety features have been implemented
according to the Bluetooth SIG requirements and tested
according to those specifications.
Use of the Naída CI M90 sound processor next to or
stacked with other equipment should be avoided because
it could result in improper operation (e.g., poor sound
quality, cutouts in sound, delay in response when making
adjustments, or loss of lock). If such use is necessary,
the Naída CI M90 and the other equipment should be
observed to verify that they are operating normally.
Use of components, accessories, and cables other
than those specified in the Compatibility section of this
document could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this
equipment and result in improper operation (e.g., poor
sound quality, cutouts in sound, delay in response when
making adjustments, or loss of lock).
50