SSCOR S-SCORT VX2 Manual De Instrucciones De Funcionamiento Y Mantenimiento página 10

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

WARNING: DO NOT ATTEMPT TO SERVICE THIS EQUIPMENT IF YOU ARE NOT
To Remove Canister Holder:
N
TO REMOVE CANISTER HOLDER:
Remove one 4" screw and two set screws on the canister bracket. Lift canister holder up and out of
the main chassis.
OPEN/ CLOSE
Clamshell design enables access to components. All 6- 32 phillips head fastening screws located in
back of the unit.
REMOVE MAIN CHASSIS SCREWS
(A)
1 screw under the canister holder.
(B)
2 screws in the back of the handle.
(B)
3 screws in the back of the main chassis.
(C)
It should not be necessary to open the exhaust door unless fluids have entered the system.
Slowly pull the unit apart at the handle to observe internal
components.
(E)
PC BOARD- ELECTRICAL CIRCUITS
Do not attempt to service the PC Board. Return to SSCOR for service.
(I)
Battery Bracket.
(J)
Battery, 12V, Sealed Lead Acid.
(K)
Wiring connection to the control panel.
(L)
Vacuum pump. Return to SSCOR for service.
(M)
Exhaust barb
(N)
Vacuum barb
Internal Access for Electrical and Pneumatics
S- SCORT VX2
A BIOMEDICAL TECHNICIAN
H
K
L
M
J
10
E
I

Publicidad

Tabla de contenido
loading

Tabla de contenido