Setting up
To set up the HEINE Retinoscope, screw the instrument head into the HEINE battery
handle or into the HEINE rechargeable handle.
Verify that the lamp voltage complies with the supply voltage of the handle. The coloured
marking on the bottom of the lamp shows you the lamp voltage:
White ring = HEINE XHL
only for use with the HEINE Battery handle
Red ring = HEINE XHL
only for use with the HEINE Rechargeable Handle
Operation
Two different-length brow rests (5) and (6) are included with the instrument. They can be
removed and interchanged according to user preference.
The control ring (3) is used to rotate the streak image through 360°. The lowest position
gives a divergent beam, the highest position focuses the streak at a distance of ca. 25 cm
in front of the retinoscope.
ParaStop: The slide control (4) limits the upward travel of the control ring to give a parallel
beam of light at the top of its travel. To operate, press the button in and down. To disengage,
push either the button or the control ring (3) upwards.
To convert the Streak Retinoscope to a Spot Retinoscope simply change the streak bulb
for a spot bulb.
HEINE Retinoscopes are intended for a brief examination of less than 6 minutes with
a 15 minutes break until the next application.
The setup and operation of the HEINE handles are described in a separate instruction
document.
Fixation Cards
The HEINE Fixation Cards for dynamic retinoscopy are attached to the retinoscope by
means of a special mount.
• Slide the mount from the patient's side into the lateral grooves (A).
• Select a fixation card and hook it on from above (B).
When not using fixation cards the mount can be left on the retinoscope or canbe removed
by pulling it off.
Hygienic Reprocessing
Instructions on hygienic reprocessing must be adhered to, based on national standards,
laws and guidelines.
Classification according to KRINKO: non-critical
Spaulding Classification USA: non-critical
Allow the device to cool down before reprocessing.
When contamination is suspected, a hygienic reprocessing has to be carried out
immediately or at least once a work day.
The described cleaning and disinfection measures do not replace the specific rules
applicable for the establishment.
8/48
2.5V bulb
®
3.5V bulb
®
med 1114
2014-10-15