Caring For Your Test Strips - ForaCare Pro Voice V9 Manual Del Propietário

Sistema de vigilancia de glucosa sanguinea
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Meter Storage
● Storage conditions: -4°F to 140°F (-20°C to 60°C), below 95% relative
humidity.
● Always store or transport the meter in its original storage case.
● Avoid dropping and heavy impact.
● Avoid direct sunlight and high humidity.
Cleaning and Disinfection of Your Lancing Device
For home users, the lancing device is reusable. Please disinfect your
lancing device regularly.
● To clean the lancing device, use a disinfecting wipe to clean the lancing
device.
● Disinfect the cap by placing it in 70% alcohol for 10 minutes and then
allow it to air dry.
● Do NOT place the device in a dishwasher or use detergents.

Caring for Your Test Strips

● Storage conditions: 39.2°F to 104°F (4°C to 40°C), and 10% to 85%
relative humidity. Do NOT freeze.
● Store your test strips in their original vial only. Do not transfer to
another container.
● Store test strip packages in a cool dry place. Keep away from direct
sunlight and heat.
● After removing a test strip from the vial, immediately close the vial cap
tightly.
● Touch the test strip with clean and dry hands.
● Use each test strip immediately after removing it from the vial.
● Write the opening date on the vial label when you first opened it.
Discard remaining test strips after 3 months.
● Do not use test strips beyond the expiration date. This may cause
inaccurate results.
32

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido