Solar Installation PV100
6 Electrical Specifications
6.1 Panel Voltage Measurement (Panel Terminals)
Display Range
Measuring Range
Resolution
Accuracy
6.2 Short Circuit Current Measurement
Display Range
Measuring Range
Resolution
Accuracy
6.3 Earth Continuity
Test Voltage Open Circuit
Test Current into 2Ω
Display Range
Measuring Range (EN 61557-4)
Resolution
Accuracy
Number of repeat tests as per
IEC61557-4
6.4 Insulation Resistance
Test Voltage Specification
Test Voltage @ 1mA
Test Current Short Circuit
Display Range
Measuring Range (EN 61557-2)
Resolution
Accuracy
Number of repeat tests as per
IEC61557-2
6.5 Rpe Voltage Measurement (4mm Terminals)
Display Range
Voltage Measuring Range
Resolution
Accuracy
6.6 Operating Current (via AC/DC Current Clamp)
Display Range
Current Measuring Range
Resolution
Accuracy
0.0V – 1000V
5.0V – 1000V
0.1V
±(0.5% + 2 digits)
0.00A – 9.99A
0.50A – 9.99A
0.01 A
±(1% + 2 digits)
>4V
>200mA
0.00Ω - 199Ω
0.01Ω –199Ω
0.01Ω maximum
±(2% + 5 digits)
Approx 4000
-0% +20% (open circuit)
>1mA into U
x (1000Ω/V)
N
<2mA
0.05MΩ - 199MΩ
0.05MΩ – 199MΩ
0.01MΩ maximum
0.05MΩ – 100MΩ ±(5% + 5 digits)
101MΩ – 199MΩ ±(10% + 5 digits)
Approx 3000
5V – 440V
5V – 440V dc
5V – 440V ac 50-60Hz
1V
±(5% + 2 digits)
0.1A – 40.0A
0.1A – 40.0A dc
0.1A – 40.0A ac 50-60Hz
0.1A
±(5% + 2 digits)
- 15 -
Operating Instructions