Control Solution Test
Chapter 5
Control solution is used to check the performance of the kit.
The meter kit should be check:
• When the meter and strip do not work properly.
• When the test result is unusual or inconsistent.
Control solution range is shown as follow:
Please check your strip vial label for exact range.
Control Range LV1
98-164
98-164
mg/dL
5.4-9.1
mmol/L
Storage
5.4-9.1
General Life Biotechnology Co.,Ltd .
5F., No.240, Shinshu Rd., Shin Juang Dist.,
New Taipei City 242, Taiwa n
Perform a Control Test:
Insert an Unused Strip
Press "M" Button (3 secs)
(Shown CL on Screen)
Shake the Control Solution Well
Drops
Put One Drop onto a Clean Surface
Sample Inlet
Test Result Display after Count Down
Range Listed on Test Strip Vial
Note:
• Do not reuse the test strip.
• Marked open date on new open control solution.
• Do not use the meter if the control test is out of range.
• If the control test keeps result wrongly, please contact an authorized distributor.
• Control solutions are not included. Please contact an authorized distributor for purchasing.