Guidance and manufacturer's declaration - Enclosure Port
The [MDS3001/MDS3001PLUS/MDS3001U] is intended for use in the electromagnetic environment
specified below. The customer or the user of the [MDS3001/MDS3001PLUS/MDS3001U] should assure
that it is used in such an environment.
Phenomenon
Electrostatic
Discharge
Radiated RF EM field
Proximity fields from RF
wireless communications
equipment
Rated power frequency
magnetic fields
Basic EMC
standard
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
-34-
Immunity test levels
Home Healthcare Environment
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
10V/m
80MHz-2.7GHz
80% AM at 1kHz
Refer to Proximity fields from RF wireless
communications equipment
30A/m
50Hz or 60Hz