Risk of injury due to electric shock, fire, ignition, explosion, leaking or bursting of
the batteries!
To avoid injuries or property damage, e.g. due to electric shock, fire, ignition, explosion,
leaking or bursting:
Use only the batteries and charging units provided in the delivery for the headlight
system purchased.
Charge and use batteries only in ambient temperatures from 0°C to +30°C (+32°F to
+86°F).
Use the batteries only at a relative humidity of 45–85%.
Store batteries only at ambient temperatures from 0°C to +25°C (+32°F to +77°F).
Do not open the batteries, throw them into fire, insert them inverted, use them together
with used batteries or batteries of another type, short-circuit them, disassemble them,
heat them to over 100°C (122°F) or dispose them into waste incineration plants.
Do not touch damaged/leaking batteries. Immediately and properly dispose of the
batteries, observing the handling and safety measures for LiPo batteries.
4.3

Charging Unit

Use the charging unit only in dry environments in a closed area. Protect from rain, snow,
inflammable materials, excessive heat and/or moisture.
Do not connect the charging unit to an extension cable.
Do not excessively bend the mains cable or place any heavy objects on it; do not pull, drag
or strain the cable.
Do not use the charging unit if the cable has been severed or the batteries and/or the
4-chamber charging unit have been dropped or damaged.
Do not used damaged charging units; do not disassemble the charging unit or modify any
parts of the charging unit.
Do not use unapproved power supply units, and do not charge damaged batteries.
Do not bring metallic objects into contact with the metal parts of the DC connectors on the
charging unit or the battery contacts.
28