Troubleshooting - VITLAB maneus Instrucciones De Manejo

Auxiliar de pipeteado / pipette controller
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

12. Troubleshooting

Trouble
Suction capacity
impaired
Pipette drips
Pipette not held
properly
Compressed
suction bellows
(1) take in air
automatically
Note: If recommendations in the Troubleshooting do not resolve
current problems, contact your supplier.
32
Possible Cause
– membrane filter (5) dirty
or
– valve unit (3) damaged
– pipette not inserted far
enough or
– membrane filter (5) not
properly positioned or
– valve unit (3) damaged
– adapter (6)
damaged
– suction (1) bellows
or
– valve unit (3) damaged
Action to be taken
– replace membrane
filter (5)
– replace valve unit (3)
– carefully push pipet
further into securing
sleeve or
– insert membrane filter
(5) or
– replace valve unit (3)
– replace adapter (6)
– replace suction
bellows (1) or
– replace the valve unit
(3)