CALIBRATION
1. Press
2. It displays "d *", * means division value, you can choose 1, 2, 5, 10, 20,
50 and then press
3. It displays "P *", * means decimal points of division value, choose 0 for
no decimal or 1 means 1 decimal; then press
4. It displays Max capacity like "05000", input Max Cap you want and press
.
5. It displays "noload", wait until stable light on and press
6. It displays f.S., load weights onto the scale (like 3000kg), input value of
weights like "03000", wait until the stable light on, then press
7. Finished.
12
, input password "13386" and press
.
.
.
.
.