Annex A

10.1 Electromagnetic Compatibility

The product is suitable for use in an specific electromagnetic environment.
The customer and/or the user of the product should assure that it is used in an electromagnetic environment as described below.
Emission Test
RF-emission CISPR 11
RF-emission CISPR 11
Harmonic emissions IEC 6100-3-2
Voltage fluctuations/
flicker emissions IEC 61000-3-3

10.2 Electromagnetic Immunity

The product is suitable for use in a specific electromagnetic environment.
The customer and/or the user of the product should assure that it is used in an electromagnetic environment as described below.
Emission Test
Electrostatic discharge(ESD)
IEC61000-4-2
Electrical fast transient/bursts
IEC61000-4-4
Surge
IEC61000-4-5
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Compliance
The product use RF energy only for its internal function.
Group 1
Therefore, its RF emissions are very low and not likely to cause any interference in nearby electronic equipment.
Class A
Class A
The product is suitable for use in all establishments, including domestic establishments and those directly
connected to the public low-voltage power supply network that supplies buildings used for domestic purpose.
Complies
IEC 60601- Level
Compliance Level
± 6kV contact
± 6kV contact
± 8kV air
± 8kV air
± 6kV contact
± 6kV contact
± 8kV air
± 8kV air
± 6kV contact
± 6kV contact
± 8kV air
± 8kV air
<5% UT (>95% dip in UT)
<5% UT (>95% dip in UT)
for 0.5 cycle 40% UT
for 0.5 cycle 40% UT
(60% dip in UT)
(60% dip in UT)
for 5 cycles 70% UT
for 5 cycles 70% UT
(30% DIP IN UT)
(30% DIP IN UT)
for 25 cycles <5% UT
for 25 cycles <5% UT
(>95% dip in UT) for 5 sec
(>95% dip in UT) for 5 sec
3A/m
Electromagnetic Environment Guidance
Floor should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the relative humidity should be at least 30 %
Mains power quality should be that of a typical commercial and/or hospital environment
Mains power quality should be that of a typical commercial and/or hospital environment
Mains power quality should be that of a typical commercial and/or hospital environment.
If the user of the product requires continued operation during power mains interruptions,
it is recommended that the product be powered from an uninterruptible power supply or
a batter
Power frequency magnetic fields should be at levels characteristic of a typical location in
3A/m
a typical commercial or hospital environment.
91
Electromagnetic Environment Guidance

Publicidad

Tabla de contenido
loading

Tabla de contenido