View: Press
20 measured/calculated values will be
displayed in a reverse order.
Trouble shooting
All errors or failures will be shown as codes.
The following table explains the meaning of
codes and solutions.
Code
Calculation
204
Excessive
208
Low battery
220
capacity
Temperature
252
too high
Temperature
253
Received light
too weak or
255
measurement
time to long
Received
256
signal too
Outside of
the range of
261
measure-
button in a row, the last
Cause
error
current
too low
strong
ment
Solution
Refer to user
manual, repeat the
procedures.
Please contact
your distributor
Replace the battery
Let the instrument
cool down
to operating
temperature at 0℃
~ 40℃
Warm up the
instrument
to operating
temperature
Use target plate
or change a good
reflective surface.
Target too
reflective, user
target plate or do
not aim at strong
light objective.
Select the
measurement
distance within
the range of
measurement.