1 Introduction

1.1

Instruction and Operation Manual for ATEX

This section covers only the ATEX-relevant aspects of the product.
Refer to SLS510 ATEX certificate KEMA 13ATEX0081 or KEMA 13ATEX0082. When load cells are considered in type of protection
non-sparking Ex nA, they need to be fitted in an enclosure that complies with the equipment's of clause 6 of EN60079-15. With a
permanent marker place a mark (√) in the box on the load cell label to indicate the applicable protection (KEMA 13ATEX0081 or KEMA
13ATEX1082). Once selected it may not be changed.
FM 09ATEX0048X1
II 1 G Ex ia IIC T4 Ta = -20°C to +40°C IP67
II 1 D Ex iaD T73°C IP67
Ui = 20V ; li = 600mA ; Pi = 6W
Ci = 12 nF ; Li = 40 μH
EN 60079-0: + A11 2013
EN 60079-11 2012
EN 60079-26 2007
EN 60529 : + A2 2013

1.2 Year of Manufacture

Not applicable

2 Function of Equipment

The SLS510 can be used as category 1 or 3 equipment for hazardous Gas and hazardous Dust environments (Zone 0, 1, 2, 20, 21 and 22).

2.1 Details on Temperature Class / Coding

The following table shows the relationship between maximum total power Pi and maximum ambient temperature.
Temperature class/coding
T4 (gas)
T73°C (dust)

2.2 Connection

Color Code:
GREEN
BLACK
WHITE
RED
YELLOW
The intrinsically safe circuit including the load cells must be built up with approved safety barriers or switch amplifiers matching the
connected weighing indicator.

2.3 Information for Connections

a. The resistance between all safety ground connections and the system grounding electrode must not exceed 1 ohm.
b. If applying as intrinsically safe, the load cells must be interfaced with an approved safety barrier matching the entity parameter of the
load cells.
c. If applying as intrinsically safe, equipment connected to the safe side of the barrier shall not be powered by or generate more than
250VAC.
64068638 SLS510 Load Cell
FM 09ATEX0049X1
II 3 G Ex nL IIC T4 Ta = -20°C to +40°C IP67
Ui = 20V ; li = 600mA ; Pi = 6W
Ci = 12 nF ; Li = 40 μH
EN 60079-0: + A11 2013
EN 60079-11 2012
EN 60529: + A2 2013
Ui = 20 V, Pi = 6W
- 20°C ≤ Ta ≤ 40°C
- 20°C ≤ Ta ≤ 40°C
+ Excitation
- Excitation
+ Signal
- Signal
Shield
3