Konica Minolta PULSOX-1 Manual Del Usuario página 18

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Error
Messages
The pulse is weak and
pulse signals required
for measurement are
Pulse
not obtained.
rate
The stated accuracy
readout
of pulse rate measure-
flashing
ments is invalid.
The battery is low.
Lit
The battery runs out.
Flashing for
about 2 sec.
then power off
The measuring cir-
cuitry of the instrument
has failed. Measure-
ment is disabled.
Error
indication
*1 Contact the nearest authorized service facility.
E6
Cause
Solution
Properly insert a fingertip into
1
the finger holder.
This is likely caused by
2
insufficient blood circulation
to the fingertip. Rub or warm
the fingertip or use another
finger.
The battery will run out in less
than 6 hours. Replace the bat-
tery as early as possible.
Replace the battery.
If this message persists after the
battery is replaced, the instru-
ment has likely failed. *1

Publicidad

Tabla de contenido
loading

Tabla de contenido