4. Safety Precautions
To ensure correct use of this instrument, read the following points
carefully and adhere to them. After you have read this manual,
keep it in a safe place where it can be referred to anytime a ques-
tion arises.
Warning
(Failure to adhere to the following points may result in death or serious
injury.)
This device is designed to measure the oxygen saturation level (SpO
•
and the pulse rate. It should not be used as a warning device to detect
cessation of breathing or irregular breathing during sleep.
This instrument has no alarm function. Do not use this instrument if an
•
alarm function is necessary.
Do not use the instrument in places where flammable or combustible
•
gases including anesthetic gases, high- purity oxygen, or gasoline va-
por are present. Doing so may cause a fire.
Do not disassemble or modify the instrument. Doing so may cause a
•
fire or electric shock.
The instrument should not be operated if it is damaged, or smoke or
•
odd smells occur. Doing so may result in a fire. In such situations, turn
OFF the power immediately, remove the battery, and contact the near-
est authorized service facility.
Do not put the battery on a fire, short-circuit it, heat it or disassemble
•
it. Doing so may cause explosion or heat generation, resulting in fire
or injury.
Do not use the instrument in conjunction with hyperbaric oxygen
•
equipment. Doing so may cause explosion or fire.
Do not use the instrument in conjunction with an MRI system. Doing
•
so may cause partial heating due to induced electromotive force, re-
sulting in burn.
)