English
Technical specifications
Earth resistance
measurement
Measurement of voltage
to earth
Insulation resistance
measurement
AC voltage measurement
General specifications
Electrical safety
Power supply
Safety fuses
Mechanical specifications
Dimensions
Mass
Accessories
Accessories supplied
with the instrumen
Analog earth & insulation tester
3 automatic measurement ranges: 0...12 Ω / 0...120 Ω / 0...1200 Ω
constant measurement current: approx. 2 mA / 820 Hz
accuracy: ± (3% + 1 mm) end of scale
measurement range: 0...30 VAC / 40...500 Hz
accuracy: ± (2.5% + 1 mm) end of scale
2 measurement ranges:
0...100 MΩ at 250 VDC
0...200 MΩ at 500 VDC
accuracy of test voltage: nominal output voltage ± 10%
short-circuit current: 2 mA DC
accuracy of insulation resistance: ± 5% of length of scale
measurement range 0...600 VAC
accuracy: ± 2.5% of full scale
line frequency range: 40 ...1000 Hz
EN 61010-1, Cat.III, 600 V
8 x 1.5 V batteries (AA or equivalent)
1 A / 250 V 0.2 Ω
250 x 190 x 110 mm
~1,500 g with batteries (exclusive of accessories)
EIT 810.............................................................................P06.2347.02
Supplied with:
- 1 carrying strap
- 1 earth kit comprising 2 rods, 3 red/yellow/green cables for the 3P method,
and - 1 double cable for the 2P method,
- 2 red and black leads with alligator clip ends for insulation measurements,
- 8 x 1.5 V batteries,
- 1 spare fuse,
- 1 user's manual in five languages