ELECTROmAGNETIC COmPATIBILITY (EmC)
EN
Recommended separation distances between portable and mobile RF communications
Cryo Cuff IC is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of Cryo Cuff IC can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and Cryo Cuff IC as recommended below, according to the maximum output power of
Rated Maximum
Output Power of
Transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
equipment and the Cold Therapy Unit
the communications equipment.
SEPARATION DISTANCE ACCORDING TO
FREQUENCY OF TRANSMITTER
150 kHz to 80 MHz
d = [3.5/V
]√P
1
0.12
0.38
1..2
3.8
12
m
80 MHz to 800 MHz
d = [3.5/E
]√P
1
0.12
0.38
1.2
3.8
12
12
800 MHz to 2.5 GHz
d = [7/E
]√P
1
0.23
0.73
2.3
7.3
23