INDICATOR
NAME
Lc
Password key
ndA
Temp alarm
mode
ndt
Timer mode
Hn
Timer unit
Timer end
EH
mode
Auxiliary
oPn
parameter
Maximum
nP
power
Co
Off point
Minimum
SPL
set point
Maximum set
SPH
point
7. Fault analysis
FAILURE
No power supply
No temperature rise
There is a big
difference between
display temp. and
actual temp.
INSTRUCTIONS OF THE FUNCTION
The password to modify the parameters in this table is
Lc = 9
0: With over-temp alarm only.
1: With over-temp alarm and under-temp alarm at the
same time.
0: No timer function.
1: With timer function, the lower window displays the
running time after the measured temperature reaches to
the set value.
2: With timer function, the lower window always displays
the setting value.
0: Minutes
1: Hours
0: Continue to maintain the constant temperature when
the running time is over.
1: Stop the temperature control when the running time is
over.
0: ON
1: OFF
Percentage of maximum power heating output.
If SV > (SP + Co), stop the heating output forcibly.
Adjustment of the minimum temperature.
Adjustment of the maximum temperature.
CAUSE
Bad contact between plug and socket
The fuse is burnt
The temperature controller is broken
The sensor is broken
The set temperature is lower than water
temperature
The heating pipe is burnt
The temperature controller
The temperature sensor is broken.
Replace the plug or socket tube
Replace the fuse with one with the
same specifications
Replace the instrument
Replace the sensor
Reset the temperature
Replace the heating pipe
Replace the temperature controller
Replace the temperature sensor
8
(RANGE)
DEFAULT VALUE
0
(0~1)
0
(0~2)
1
(0~1)
0
(0~1)
0
(0~1)
0
(0~100%)
100
(0~100.0℃)
50.0
(0~0)
0
(0~400.0)
300.0
HANDLING METHOD