5.6. Temperature measurement
To measuring temperature shuld be use the K-type probe: P3400.
■
Set the rotary switch to the "TEMP" range. In this time, The environment temperature
value displayed on the LCD.
■
nsert the K-type probe to the two "COM" and "TEMP" terminals. The two "+" and "-"
pins of temperature probe (P3400) must be direct at the "COM" & "
respectively.
■
he measured temperature value will be displayed on the LCD.
6. Maintenance
6.1. Replacing the battery
When meter display
■
Disconnect and remove all test probes from any live source and meter.
■
Open the battery cover on the bottom case by screwdriver.
■
Remove old battery and snap new one into battery holder
6.2. Fuse replacement
Replacing the defective fuse should the done according to the following procedure.
■
To avoid electrical shock, remove the test lead and any input signal before opening the
bottom case.
■
Open the botton case and then remove the defective fuse and insert a new fuse of the
same size and rating.
■
Replace the bottom case and reinstall all the screw.
6.3. Cleaning and Decontamination
The meter can be cleaned with soft clean cloth to remove any oil, grease or grim. Do not
use liquid solvent or detergent.
the battery must be replace to maintain normal operation.