Installation instruction
Equipotential bonding
Also on the rear side of the UV Plus there
is a stud for the equipotential bonding.
Integrate the UV Plus with a suitable cable
into the equipotential bonding system.
Observe on Combi-Duo:
Attention
UV Plus must not be installed on Combi Duo with
caster. Danger of tipping over!
-
Connect the vent pipe of the Combi-Duo to the
connection piece of the UV Plus using hose
(item 10) and hose clamp (item 11) of the
UV Plus kit.
-
Installation of the bus cable:
There are two bus cables connected to the con-
trol pcb. Use the longer cable for the bottom unit
and the shorter for top unit. Connect the cable
as described under point 4.
7. Technical intervention
Before any intervention or maintenance work,
disconnect the appliance from the power sup-
ply.
7.1 Fan motor.
Fan motor is accessible after removing the
fat filter and pre filter.
7.2 Changing halogen bulb.
After loosen the holding frame, the halogen bulb
can be changed.
7.3 Changing differential pressure switch of pre filter
- Set adjustment disc of pressure switch to zero
position
- Switch on Ultravent and unit
- Remove fat filter. On the pre filter cover
approx 2/3 of the surface of the filter with a
piece of paper.
- Rotate adjustment disc out of zero position
until LED of pre filter switches on.
- When removing the paper LED must switch of.
UV Plus V05 08/2015
8. Wiring diagram
see Page 80
9. Service parts
Description
Control pcb
Bus cable
Fuse 5 x 20 2,5A
Halogen bulb
Inner gasket for interior light
Glass pane for interior light
Gasket frame for interior light
Fat filter left
Fat filter right with lock
Adapter plate left hinged door
(Pos. 1) 61/101
Adapter plate right hinged door
pic. 9
(Pos. 1) 61/101
Adapter plate right hinged door
(Pos. 1) 62/102
Gasket adapter (Pos. 2) 61/101
Gasket adapter (Pos. 2) 62
Gasket adapter (Pos. 2) 102
Gasket adapter (Pos. 2) 201
Drain pipe (Pos. 6)
Silicone hose
Contactor K1/K2
Capacitor 3,5 uF
Capacitor 10 uF
pic. 17
Capacitor 14 uF
pic. 15
Capacitor 40 uF
LED yellow
LED red
Relais kpl. 1U 24V
Differential pressure switch
Control transformer
Voltage related parts
Extractor fan 200-240V (M1)
Extractor fan 100-127V (M1)
- 12 -
42.00.050
40.04.148
3019.0117
3024.0201
40.00.094
40.00.095
40.00.092
60.72.440
60.72.441
60.72.891
60.72.936
60.72.937
60.72.431
60.72.433
60.72.435
60.72.445
60.70.735
2110.1008
40.03.687
40.04.151
40.04.152
40.04.229
40.04.230
40.04.165
40.04.166
40.04.304
3017.1014
3037.0253
for input voltage:
60.72.439
60.72.701
Art. No.