Boeco BLC Serie Instrucciones De Manejo página 6

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

To select Unit 2
1. Turn the balance off.
2. Hold down the [ON/TARE] key until
all segments are displayed.
3. Continue to hold down the
[ON/TARE] key and press the
[UNITS] key four times.
4. Release the [ON/TARE] key.
5. Press the [ON/TARE] key three times
so the display reads Unit 2.
6. Press the [UNITS] key to scroll
through the choices of units.
7. Press the [OFF] key when the
desired unit appears.
6
All manuals and user guides at all-guides.com
C. Calibration
To calibrate the balance, tare or zero
the balance with no weight on the pan.
1. Turn the balance off.
2. Hold down the [ON/TARE] key until
all segments are displayed.
3. Continue to hold down the
[ON/TARE] key and press the
[UNITS] key three times. Release the
[ON/TARE] key.
4. After a few seconds the display
C
will show
and the value of the
required calibration weight.
5. Place the correct weight on
the balance. After several seconds
the maximum capacity will be
displayed, indicating the value has
been accepted.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Blc-500Blc-3000

Tabla de contenido