Classification
Medical class:
Degree of protection
against electric shock
Permitted current
leakage
Protection reliability
Power supply
Power consumption
Mode of operation
Protection against risk of
explosion
Protection against liquid
penetration
Degree of mobility
Type of approval
Safety test in line with
EMC test in line with
Primary fuse
Safety
Users of this form of lighting should be aware that the manufacturer of the product
can only accept liability for the lighting when maintenance, repairs or changes are
carried out by the manufacturer or by companies which the manufacturer has
approved to carry out such work, and when components that influence the safety of
the lighting are replaced with original spare parts. The lighting is not intended for use
in areas where there is a risk of explosion.
This product complies to IEC/EN 60601-1-2(2001), electromagnetic compability for
medical devices. Electromagnetic or other interference between this equipment and
other devices should not occur. In the event of interference, please move the
equipments apart from each other.
The power cable must only be replaced by a qualified electrical fitter or other
similarly qualified professional.
Any repairs of the arm must be performed by LUXO or LUXO approved personnel.
The arm contains high tensions springs.
Maintenance
Cleaning of product and replacement of light source is the only maintenance needed.
ART.: 05-900-585-00
BRU-303-0029 Rev.nr.: E
LFM
Class I (grounded)
230AC 50 Hz
Max. 95VA
Not protected
IP20
Mobile table
ENEC 17
EN 60598-2-4(1997),
EN 60598
1(2004)+A11(2000)+
A12(2002)
EN
55015(2000)+A1(2001)+A
2(2002)
No fuse
LFM Medical
Class 1
Equipment type: B Applied Part
Max 0.5 mA
Resistance in earth circuit –
earth connection 0.1
230AC 50 Hz
Max. 95VA
Continuous
Not protected
Not protected
Mobile table
MEDICAL Standard
IEC/EN 60601-1(1990) 2 Ed
+A1(1993)+A2(1995)+
A13(1996)+CORRIGENDIUM(1
994)
IEC/EN 60601-1-2(2001)
IEC 127-5X20mm, 2xT0.4AH
250V
Page 8 of 21