Rated maximum output power
of transmitter W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and people
21. TECHNICAL FEATURES
21.1 Pump
MODEL
Power supply
Fuse T1AL
Power cord
Pressure output
Air release
Cycle time
Noise
Dimensions
Weight
Environment
temperature
Environment
humidity
Atmospheric pressure
Classification
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d=1.2√ P
0.12
0.38
1.2
3.8
12
LAD462
~ 220V 50/60Hz
250V
2,20 m
40~100mmHg
6/7LPM
6 min
32-34 dB
27x13x10,5 cm
~ 1,1 Kg
Operation: +10°C/+40°C
Storage: -18°C/+43°C
Shipping: -18°C/+43°C
Operation: 10%/75% non condensing
Storage: 10%/95% non condensing
Shipping: 10%/95% non condensing
70/101,325 KPa
Class II, Type BF, IP21
Applied part: air mattress
Not suitable for use in the presence
of a flammable anesthetic mixture
(No AP or APG protection)
80 MHz to 800 MHz
d=1.2√ P
0.12
0.38
1.2
3.8
12
LAD463 - LAD464
~ 220V 50/60Hz
250V
2,20 m
40~100mmHg
6/7LPM
12 min
32-34 dB
27x13x10,5 cm
~ 1,1 Kg
Operation: +10°C/+40°C
Storage: -18°C/+43°C
Shipping: -18°C/+43°C
Operation: 10%/75%
non condensing
Storage: 10%/95% non condensing
Shipping: 10%/95% non condensing
70/101,325 KPa
Class II, Type BF, IP21
Applied part: air mattress
Not suitable for use in the presence
of a flammable anesthetic mixture
(No AP or APG protection)
13
800 MHz to 2,5 GHz
d=2.3√ P
0.23
0.73
2.3
7.3
23