It is important to clean your Nebulizer before
the first time it is used and after each use.
Remember: Disconnect the unit from the electrical
outlet and never immerse the unit in water.
Cleaning the Parts
The Nebulizer, mouthpiece, masks, clear cover
and other optional accessories should be
cleaned with a mild detergent or commercially
available disinfectant.
Thoroughly rinse each part and allow to
completely air dry before using or storing.
Rinsing Technique
(Performed after each treatment):
1) Make sure that the POWER switch has been
pressed "off" and the unit has been
disconnected from the power source.
2) Disconnect the nebulizer, air tube,
mouthpiece and/or mask.
NOTE: If removing the Nebulizer
from the stem of the
Compressor, twist the entire
Nebulizer, then lift to remove.
3) Gently twist and pull up on the
cover of the Nebulizer to open
and separate, Fig. 1.
4) Rinse the Nebulizer and
components with hot tap water.
5) Dry with clean towels or allow
to completely air dry.
6) Reassemble the Nebulizer and store the
parts in a cool, dry place.
Cleaning Technique
(Performed after the last treatment of the day
or before first time use):
1) Follow steps 1-4 under "Rinsing Technique".
2) Wash all Nebulizer parts, mouthpiece and
other optional components in warm water
and a mild detergent. Rinse in hot tap water.
NOTE: For first time cleaning or after the unit
has been stored for an extended period of
time, thoroughly clean all components. The
Nebulizer is dishwasher safe.
3) Prepare a vinegar/distilled water solution of
one part white vinegar to three parts
distilled water.
4) Submerge the rinsed parts in the
vinegar/distilled water solution for
30 minutes.
5) Follow steps 4-6 under "Rinsing Technique".
Cleaning the Main Body (Compressor)
The compressor should be cleaned with a soft
cloth moistened with water and dried with a
soft cloth.
NOTE: Any other form of cleaning or
cleaning agents may damage the finish on the
unit.
Changing the Air Filters
It is important to change the air filter
approximately every 30 days or when the air
filter turns gray.
1) Remove the air filter cover by gently
peeling back the rubber cover.
2) Discard the gray air filter.
3) Replace with a new, clean air filter.
4) Securely reattach the air filter cover to the
unit.
NOTE: Air filters cannot be cleaned or washed.
Only CompXP air filters can be used. Do not
substitute alternate material such as cotton. Do
not operate without an air filter.
Using the AC Adapter,
DC Adapter or Rechargeable
Battery Pack
Your Nebulizer operates by use of an AC
adapter, DC adapter or battery pack, all
Fig. 1
included with your Nebulizer.
NOTE: The battery pack must not be connected
to the compressor when using AC or DC
voltage.
Using the AC Adapter,
1) Insert the AC adapter plug into the
dreceptacle of the compressor, located on
the bottom of the compressor, Fig 2.
2) Following general safety precautions, plug
the adapter into a standard AC outlet.
3) Proceed to "Assembling and Operating Your
Nebulizer"
instructions.
Using the DC
Adapter,
1) Insert the DC adapter
plug into the
receptacle of the
compressor, located
on the bottom of the
compressor, Fig. 2.
4
Fig. 2