IMPORTANT! Do not disassemble. Do not modify
this instrument. It may damage the instrument.
• Do not use and store the micrometer at sites where
the temperature will change abruptly. Prior to
use, thermally stabilize the micrometer at room
temperature.
• Do not store the micrometer in a humid or dusty
environment or use where the micrometer could be
subjected to water and oil.
• Protect the micrometer from sudden impacts,
including a drop.
• Always zero the micrometer prior to measurement.
• Wipe off dust, cutting chips, and moisture from the
micrometer before and after use.
• To clean the micrometer, use a soft cloth soaked
in a diluted neutral detergent. Do not use organic
solvent (thinner, etc.). It may deform or damage the
instrument.
INSTALLING THE BATTERY
IMPORTANT! When the battery has been
installed, first press the ORIGIN button to zero
the micrometer, but do not rotate the thimble
while the preset values are being set.
(See Figure 1)
• Rotating the thimble at this time may result in the
failure of setting default settings by the electrical
unit. This will prevent obtaining a correct count.
Reinstall the battery if you moved the thimble during
this time.
ORIGIN
Figure 1
• If the battery is removed then reinstalled, the origin
(zero point) set is canceled. If this happens, set the
origin again. (Refer to Zeroing the Micrometer on
page 3)
PAGE 2 OF 9
SAFETY CONTINUED
• Do not grip the display. It may damage the
micrometer.
• The spindle is designed so that it cannot be
removed from the inner sleeve. Do not move it past
the upper limit of the measuring range, as it may
damage the micrometer.
• Do not use an electric engraving pen or other such
devices on the micrometer. It may damage the
micrometer.
• Do not charge or disassemble the battery. Doing so
may cause a short circuit.
• If the micrometer is not in use for more than three
months, remove the battery. The battery could leak
and cause damage to the micrometer.
• The warranty shall not apply if the product fails or
is damaged as a result of wear and tear, including
battery drain.
SETUP
• Use only an SR44 button-type silver oxide cell. The
battery is used only for the purpose of checking
the functions and performance of the micrometer,
therefore it may not satisfy the specified battery life.
• Carefully mount the battery cover so that the cover
is completely engaged with the hole thread and
the seal does not extend off the cover. Unless the
battery cover and the seal are mounted properly, the
micrometer may not display a correct value.
• When an abnormal display appears, such as an
error display or count failure, the battery should be
removed and then reinstalled.
• Follow the local ordinances or regulations for battery
disposal.
• Install the supplied battery. (See Figure 2)
Figure 2
Battery
0421 IH-2780
Seal